WELCOME YOU

Giá : Liên hệ

Mã sản phẩm : JW-i3000

Manufacturer: JW Medical
Made in Korea

Số lượng sản phẩm:

Configuration includes:
- Main body with 10.1 inch display: 01 pcs.
- X-ray cassette tray: 01 pc.
- Skin sensing system (Dual sensor): 01 set (02 units).
- Temperature sensor: 01 unit.
- Uncoated mattress: 01 pcs.
- Pile infusion: 01 unit.
- Hepa filter: 03 units.
Handle covers: 03 pcs.
- Casing cover incubator: 01 unit.
- Power cord: 01 pcs.
- Manual and maintenance manual: 01 set (English and Vietnamese translation)
- Process of operation: 02 sets.
Features and Specifications:
1 General Features:
• Cages are used to care for infants, premature babies.
• Microprocessor controlled to control temperature and humidity in the incubator.
• Air circulation system.
• X-rays can be given inside the incubator.
• Software that controls air temperature, skin temperature, humidity. Display data on TFT LCD screen.
• The humidifier can be removed and cleaned with water.
• The monitor can rotate in 1050 different directions.
• The incubator body is designed with grooves, which easily attach accessories such as infusion pens, device trays, and accessory baskets.
2 Specifications:
• Air velocity along the mattress is 10 cm / sec.
• Even temperature on the mattress <0.80C.
• Control mode: According to skin temperature / air temperature.
• Warm-up time: <20 minutes.
• Audible noise level <45 dBA.
• The machine uses an electrostatic filter. 99.99% filter efficiency; dust particle size: 0.3 microns.
 Dashboard:
• Display: TFT LCD touch screen, size 10.1 inch.
• A warning light is on the top of the screen.
• Display of parameters: air temperature, skin temperature, set temperature, percentage of humidity, percentage of humidity setting.
• Displays algebraic and numerical values ​​on the screen.
• Control by touch keys on the screen.
 Capture:
• Cage size: 830 x 480 x 510 mm. (Long x G x Width).
• Height of the cage against the base: 1600 mm.
• The cage is made of transparent material. Two partition walls help prevent heat loss when opening cage doors.
• Port: 4 operation ports and 2 air gates.
• Wiring: 5 ports on each side, 1 port on the top of the cage.
• The mattress is made of non-X-ray material.
• Mattress size is around 700 x 20 x 370 mm (Length x Thick x Width).
• The bearing capacity of the mattress is 12 Kg.
• X-ray cassette tray size: 255 x 305 mm. (10 x 12 inches)
• Tilt of the mattress: ± 120.
• Control the tilt of the mattress by turning the knob.
Base:
• Rectangular, 4-sides with silicone cushions for impact protection.
• Dimensions: 1060 x 660 mm (Long x Width).
• 4-inch, 4-inch foot pedal, with a separate brake for each wheel.
Storage compartment:
• Made of soft plastic. Slideway.
• Dimension: 480 x 430 x 210 mm (Long x Width x Height).
 Skin temperature:
• The incubator uses two skin temperature sensors at the same time. (Dual sensor), which helps to control the temperature of the newborn.
• Display range: 25 to 450C.
• Control range: 34 to 390C.
• Accuracy: ± 0.30C.
• Resolution: ± 0.10C.
 Air temperature in the cage:
• Display range: 5 to 500C.
• Control range: from 20 to 390C.
• Accuracy: ± 0.30C.
• Resolution: ± 0.10C.
Humidifier:
• Control mode: Automatic control.
• Range of control: from 30% to 99%. Each adjustment step increases / decreases by 1%.
• Display range: 0% to 99%.
• Capacity of water tank 1500 ml.
Safety warning function: Warning with light and sound. Warning in the following cases:
• Power failure: When power is off or wires are disconnected.
• When system error occurs.
• Heat: When the air temperature exceeds the control range of 1.50C.
• Airflow: In case the fan or motor does not work.
• When the sensor is not connected to the wire or the sensor is damaged.
• Air temperature: When the air temperature is different from the set temperature above 1.00C or below 2.00C.
• Skin temperature: When the skin temperature is different from the set temperature above 1.00C.
• Low water level: When the water level of the humidifier is lower than the humidity limit.
Infusion pills:
• Made of stainless steel.
• The pile is attached to the body of the incubator. It is possible to adjust the height of the pile by adjusting the position of the pile.
• Height: 1050 mm.
• Diameter: 15 mm.
Electricity supply:
• Loop: 220 - 230V ± 10%.
• Frequency: 50/60 Hz
• Fuse: 250V; 4A
Power Consumption:
• Resistance to fire: 450W ± 10%.
• Controller: 80W ± 10%.
• humidifier: 200W ± 10%.